PN 153-691-2 www.cascademicrotech.comI Pyramid Probes: Used Core Inspection• 1
This guide describes the visual inspection process for used Pyramid Probe cores. A Pyramid Probe core is considered
“Used” once it has touched down on a wafer. Visual inspection should be performed first visually, and then with a
microscope. The recommended procedure employs two microscopes:
This guide describes the visual inspection process for new Pyramid Probe cores. Inspection should be performed first
visually, and then with a microscope. The recommended procedure employs two microscopes:
When saving photos, be sure to include the serial number in the file name.
Use the Pyramid Probes Certificate of Conformance to verify the following:
The serial number on the core frame matches the Pyramid Probes Certificate of Conformance.
The product name on the Pyramid Probes Certificate of Conformance matches the product to be tested.
Step 1: Edge Sense Inspection
Pyramid Probes used on manual probe stations have an edge sense type mechanical switch installed. This step applies
only to manual probe stations.
Pyramid Probes used on manual probe stations have an edge sense-type mechanical switch installed. The edge sense
needle contacts a glass pad on the spring in some Pyramid Probes. When installed on larger frame cores, the edge sense
needle contacts the core spring directly. This step applies only to manual probe stations.
Low-magnification microscope
(with a large working space)
High-magnification microscopes
Commonly used for PCB inspection and
soldering stations
Used for inspecting the entire probe, core
window, components and membrane wing
Magnification range should be approximately
7x to 90x
Used for inspecting the probe face, probe tips
and beams
Recommended microscope objectives include
5x, 20x, and 50x
Inspection Area Criteria
Edge Sense
(if glass pad is
installed)
Are there cracks in the glass pad?
Contact local support for instructions.
Is the glass pad securely in place?
Contact local support for instructions.
Is the needle touching the glass pad, and roughly centered?
Contact local support for instructions.
Quick Reference Guide
Pyramid Probes: Used Core Inspection
Vue de la page 0
1 2 3 4 5 6 7 8

Résumé du contenu

Page 1 - Step 1: Edge Sense Inspection

PN 153-691-2 www.cascademicrotech.comI Pyramid Probes: Used Core Inspection• 1This guide describes the visual inspection process for used Pyramid Prob

Page 2

PN 153-691-2 www.cascademicrotech.comI Pyramid Probes: Used Core Inspection• 2Step 2: Low Magnification Microscope InspectionInspection Area Criteria

Page 3 - Components

PN 153-691-2 www.cascademicrotech.comI Pyramid Probes: Used Core Inspection• 3Step 3: High Magnification Microscope InspectionInspect Core Using the 5

Page 4

PN 153-691-2 www.cascademicrotech.comI Pyramid Probes: Used Core Inspection• 42. Inspect the core using the 5x objective.Inspection Area Criteria Prob

Page 5

PN 153-691-2 www.cascademicrotech.comI Pyramid Probes: Used Core Inspection• 53. Inspect the core using the 20x objective.Inspection Area Criteria Tip

Page 6

PN 153-691-2 www.cascademicrotech.comI Pyramid Probes: Used Core Inspection• 6Middle Layer PolyimideAre there any cracks in the Middle Layer Polyimide

Page 7

PN 153-691-2 www.cascademicrotech.comI Pyramid Probes: Used Core Inspection• 74. Inspect the core using the 50x objective. The primary goal at this ma

Page 8 - Example Images

PN 153-691-2 www.cascademicrotech.comI Pyramid Probes: Used Core Inspection• 8Example Images Low Magnification Scope High Magnification ScopeEntire Pr